Integral steam jacket;
Computerized command of cycles;
Validated cycles for typical hospitalary loads;
Counting of sterilizations performed;
Real-time printing of cycle parameters;
Safety-stop system for door movement in case of obstruction;
Safety system against door opening in case of power failure;
Incorporated energy and water saving systems;
Extended listing with errors associated to eventual cycle-specific failures;
Chamber electrolytic polishing;
Chamber and jacket in AISI 316L stainless steel;
Equipment fully built in stainless steel, free of ferrous materials
Optionally supplied with 1 or 2 sliding doors;
Accurate pressure and temperature sensors.
STERI 21
They are steam sterilizers designed for hospitals and small to medium-sized laboratories. The sterilization chamber is rectangular and made of stainless steel AISI 316 L. The doors are automatic vertical sliding and made of stainless steel AISI 316 L. The jacket is made of stainless steel AISI 316 L. Steri 21 steam sterilizers are controlled by electric valves and have motorized ports, so steam sterilizers do not need compressed air. Front or side service access.
Outdoor cargo car;
Rolling load platform;
Cargo baskets;
Silent air compressor;
Water softener.
Chamber pre-heating | 134 ºC |
Wrapped textiles | 134 ºC |
Wrapped instruments | 134 ºC |
Instruments in containers | 134 ºC |
Sensitive materials | 121 ºC |
Prions or Quick | 134 ºC |
Bowie & Dick | 134 ºC |
Vacuum |
Network | Features |
ELECTRIC | 3 ~ + N + G 400 V / 230 V 50 Hz 26, 32, 38 kW |
WATER | Maximum pressure 6 Bar Minimum pressure 3 Bar Connection 3/4” BSP |
PNEUMATIC | Maximum pressure 7,5 Bar Minimum pressure 4 Bar Ligação 3/8” male connection |
DRAIN | Material resistant to 100 ºC Connection 2” BS |
TECHNICAL CHARACTERISTICS (01 a 03 STU)
MODEL | CHAMBER DIMENSIONS (LxAxP) cm |
VOLUME liters |
OUTDOOR DIMENSIONS(LxAxP) cm |
40.40.66 | 40x40x66 | 105 | 70x183x98 |
50.50.66 | 50x50x66 | 165 | 80x183x98 |
50.50.100 | 50x50x100 | 250 | 80x183x132 |